| ● | Standar | | | | | | | : | Class 1 or 2 X group according to IEC 61672:2002 and | | | | | | |
| | | | | | | | | | | | class 1 or 2 according to IEC 60651:2001 and IEC 60804:2000 | | | | |
| | | | | | | | | | | | type 1 or 2 according to ANSI S1.4-1983 and S1.43-1997 | | | | | | |
| ● | Microphone | | | | | : | 1/2″ UC52 condenser type, pre-polarized, for free field. | | | | | | |
| ● | Dynamic range | | | | : | 30 dBA ÷ 143 dB Peak | | | | | | | | | | | | | | | | | |
| ● | Linear Field | | | | | : | 80 dB | | | | | | | | | | | | | | | | | | | | | | |
| ● | Acoustic Parameters | | : | Spl, Leq, SEL, LEP,d, Lmax, Lmin, Lpk, Dose, Ln | | | | | | | | |
| ● | Frequency Weights | | | : | simultaneous A, C, Z (only C and Z for Lpk) | | | | | | | | | | |
| ● | Temporal Weights | | | : | simultaneous FAST, SLOW, IMPULSE | | | | | | | | | | | | |
| ● | Integration | | | | | | : | from 1s to 99 hours with erasing function (Back-Erase) | | | | | | |
| ● | Statistical Analysis | | | : | It displays up to 3 percentile levels, from L1 to L99 | | | | | | | |
| | | | | | | | | | | | Probability distribution and percentile level calculation from L1 to L99 | |
| | | | | | | | | | | | (Option 2 “Data Logger” and option 5 “Advanced Analyzer”) | | | | |
| | | | | | | | | | | | › | Parameter: LFp, Leq, Lpk weighted A, C o Z (solo C o Z per Lpk) | | |
| | | | | | | | | | | | › | Sampling frequency: 8 samples/second | | | | | | | | | | | |
| | | | | | | | | | | | › | Classification: Classes of 0.5 dB | | | | | | | | | | | | | |
| ● | Analysis of Events (Option 2 Data | | : | › | Calculation of 5 freely-programmable event parameters | |
| | Logger” and option 5 “Advanced | | | › | Calculation of statistical levels from L1 to L99 | | | | |
| | Analyzer” ) | | | | | | | | | | | › | Event identification trigger with programmable threshold and duration
| | | | | | | | | | | | | | | | | filter | | | | | | | | | | | | | | | | | | |
| | | | | | | | | | | | | | | | › | Manual trigger | | | | | | | | | | | | | | | |
| ● | Data Logger | | | | | | | | | : | 1 sampling from 1/8 s to 1 hour and 3 profiles with 2 samples/second
| ● | Display | | | | | | | | | | | : | 3 parameters in numeric format, Backlit LCD (“LCD” option), |
| | | | | | | | | | | | | | | | Profile LAFp with 8 samples/second, Graph of sound level probability
| | | | | | | | | | | | | | | | distribution, Graph of percentile levels from L1 to L99 | | |
| ● | Memory | | | | | | | | | | | : | Internal memory equal to 4 MB | | | | | | | | | | |
| | | | | | | | | | | | | | | | External, via the HD2110MC memory card interface, using MMC or SD
| | | | | | | | | | | | | | | | cards up to 2 GB | | | | | | | | | | | | | | | |
| ● | Input/Output | | | | | | | | | : | RS232 serial and USB interfaces, AC output (LINE), DC output |
| ● | PC Programs | | | | | | | | | : | PC interface for data download, set up and instrument management.
| | | | | | | | | | | | | | | | Licensed software modules to be enabled by hardware key. | |
| ● | Operating conditions | | | | | | : | Working temperature ‑10÷50°C, 25÷90%RH (without condensation),
| | | | | | | | | | | | | | | | 65÷108kPa. Protection degree: IP64 | | | | | | | | |
| ● | Power Supply | | | | | | | | | : | 4 x AA batteries or external 9÷12Vdc 300mA | | | | | | |
| ● | Dimension and weight | | | | | | : | 445x100x50 mm equipped with preamplifier, 740 g (with batteries).
| ● | Buku Manual Operasional | | | | : | Bahasa Inggris dan Bahasa Indonesia | | | | | | | | |
| ● | Kartu Garansi | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| ● | Environtmental and Health Integration System (EHIS)
| • | Login user admin sebagai induk Instansi dapat menambah login untuk anggota dengan jumlah tak terbatas,
| | sebagai upaya pemantauan kegiatan pengujian dilapangan.
| • | Laporan yang tersimpan setiap instansi yang ditampilkan tanggal, nama alat, propinsi dan tempat pengujian